Specifications

Luminescence 96

General

Type of product

Single-mode Luminometer

Detection method

Luminescence

Detection mode

Endpoint, kinetic

Microplates types

96-well microplates, white or black

Readout

Glow, Flash (Development times > 5 seconds, manual addition of reagents)

Readout orientation

Top and bottom readout

Measurement

Detector

96 Silicon Photomultiplier

Sensitivity

100 fmol ATP/well (preliminary result)

Linear dynamic range

8 decades

Cross talk

0.7 x 10⁻⁶ RLU

Well-to-well variability

< 5 %

Data output

USB 2.0 interface with PC or WLAN

Temperature for measurement

15–40 °C

Physical characteristics

Dimensions

9,8 x 14,0 x 4,2 cm (W x L x H)

Weight

570 g

Power supply

USB connection 5 V

Power consumption

2.5 Watts

App

System Requirements

Microsoft Windows

Windows 10 (1809 or later)

macOS

11 Big Suror later

App features

Measurement modes

Endpoint, Kinetic

Evaluation methods

Quantitative, qualitative

Regression methods

Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL) and more

Export formats

PDF, CSV

Specifications

Luminescence 96 Automate

General

Type of product

Single-mode Luminometer

Detection method

Luminescence

Detection mode

Endpoint, kinetic

Microplates types

96-well microplates, white or black

Readout

Glow, Flash (Development times > 5 seconds, manual addition of reagents)

Readout orientation

Top and bottom readout

Measurement

Detector

96 Silicon Photomultiplier

Sensitivity

100 fmol ATP/well (preliminary result)

Linear dynamic range

8 decades

Cross talk

0.7 x 10⁻⁶ RLU

Well-to-well variability

< 5 %

Data output

USB 2.0 interface with PC or WLAN

Temperature for measurement

15–40 °C

Physical characteristics

Dimensions Reader

9.6 x 13.8 x 4.7 cm (W x L x H)

Dimensions SBS Adapter

8.6 x 12.8 x 1.7 cm (W x L x H)

Weight Reader

Upper part 515 g

Power supply

USB connection 5 V

Power consumption

2.5 Watts

App

System Requirements

Microsoft Windows

Windows 10 (1809 or later)

macOS

macOS 11 Big Sur

App features

Measurement modes

Endpoint, Kinetic

Evaluation methods

Quantitative, qualitative

Regression methods

Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL) and more

Export formats

PDF, CSV

Specifications

Absorbance 96 Automate

General

Detection method

Absorbance

Detection mode

Endpoint, kinetic

Microplates types

96-well microplates

Measurement

Light source

Up to 6 LEDs

Detector

96 Photodiodes

Wavelength selection

Example combinations (FWHM 10 nm): 450, 492, 562, 620 nm/405, 450, 540, 630 nm/492, 562, 605, 650 nm. Other filter combinations in the range of 400–1000 nm possible.

Photometric range

0.0–4.0 OD

Linearity

405 nm: ≤ 1.5 % from 0.0–2.0 OD; ≤ 3.0 % from 2.0–3.0 OD
≥ 450 nm: ≤ 1.0 % from 0.0–2.0 OD; ≤ 1.5 % from 2.0–3.0 OD

Accuracy

405 nm: ≤ 1.5 % + 0.010 OD from 0.0–2.0 OD; ≤ 3.0 % + 0.010 OD from 2.0–3.0 OD
≥ 450 nm: ≤ 1.0 % + 0.010 OD from 0.0–2.0 OD; ≤ 1.5 % + 0.010 OD from 2.0–3.0 OD

Reproducibility

≤ 0.5 % + 0.005 OD from 0.0–2.0 OD; ≤ 1 % + 0.010 OD from 2.0–3.0 OD

Resolution

0.001 OD

Read time

Down to 3 sec. at single wavelength

Data output

USB 2.0 interface with PC

Physical characteristics

Dimensions Reader

9.6 x 15.5 x 5.7 cm (W x L x H)

Dimensions SBS Adapter

8.6 x 12.8 x 1.7 cm (W x L x H)

Weight Reader

Upper part 435 g, lower part 355 g

Power supply

USB connection 5 V

Power consumption

2.5 Watts

App

System Requirements

Microsoft Windows

Windows 10 (1809 or later)

macOS

macOS 11 Big Sur or later

App features

Measurement modes

Endpoint, Kinetic

Evaluation methods

Quantitative, qualitative

Regression methods

Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL) and more

Export formats

PDF, CSV

Specifications

Absorbance 96

General

Detection method

Absorbance

Detection mode

Endpoint, kinetic

Microplates types

96-well microplates

Measurement

Light source

Up to 4 LEDs

Detector

96 Photodiodes

Wavelength selection

Example combinations (FWHM 10 nm): 450, 492, 562, 620 nm/405, 450, 540, 630 nm/492, 562, 605, 650 nm. Other filter combinations in the range of 400–1000 nm possible.

Photometric range

0.0–4.0 OD

Linearity

405 nm: ≤ 1.5 % from 0.0–2.0 OD; ≤ 3.0 % from 2.0–3.0 OD
≥ 450 nm: ≤ 1.0 % from 0.0–2.0 OD; ≤ 1.5 % from 2.0–3.0 OD

Accuracy

405 nm: ≤ 1.5 % + 0.010 OD from 0.0–2.0 OD; ≤ 3.0 % + 0.010 OD from 2.0–3.0 OD
≥ 450 nm: ≤ 1.0 % + 0.010 OD from 0.0–2.0 OD; ≤ 1.5 % + 0.010 OD from 2.0–3.0 OD

Reproducibility

≤ 0.5 % + 0.005 OD from 0.0–2.0 OD; ≤ 1 % + 0.010 OD from 2.0–3.0 OD

Resolution

0.001 OD

Read time

Down to 3 sec. at single wavelength

Data output

USB 2.0 interface with PC

Physical characteristics

Dimensions

9.6 x 15.4 x 5.5 cm (W x L x H)

Weight

900 g

Power supply

USB connection 5 V

Power consumption

2.5 Watts

App

System Requirements

Microsoft Windows

Windows 10 (1809 or later)

macOS

macOS 11 Big Sur or later

App features

Measurement modes

Endpoint, kinetic

Evaluation methods

Quantitative, Qualitative

Regression methods

Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL)

Export formats

PDF, CSV

Specifications

Absorbance One

General

Type of product

Single beam, single wavelength cuvette photometer

Detection method

Absorbance

Detection mode

Endpoint

Cuvette types

Micro, semi-micro and macro cuvettes

Measurement

Light source

1 x LED

Wavelength

UV: 280, 340 nm/VIS: 400–1000 nm

Measurement range

0–5.0 OD

Resolution

0.001 OD

Accuracy

≤ 0.5 % + 0.005 OD from 0.0–3.0 OD

Reproducibility

≤ 0.5 % + 0.005 OD from 0.0–3.0 OD

Linearity

≤ 0.5 % from 0.0–3.0 OD

Temperature for measurement

5–50 °C

Physical characteristics

Power supply

USB connection 5 V

Dimensions

33 mm x Ø 64 mm

Weight

150 g

App

Operating systems

Windows 10
macOS 11
Android 5