Cart
Your cart is empty
Single-mode Luminometer
Luminescence
Endpoint, kinetic
96-well microplates, white or black
Glow, Flash (Development times > 5 seconds, manual addition of reagents)
Top and bottom readout
96 Silicon Photomultiplier
100 fmol ATP/well (preliminary result)
8 decades
0.7 x 10⁻⁶ RLU
< 5 %
USB 2.0 interface with PC or WLAN
15–40 °C
9,8 x 14,0 x 4,2 cm (W x L x H)
570 g
USB connection 5 V
2.5 Watts
Windows 10 (1809 or later)
11 Big Suror later
Endpoint, Kinetic
Quantitative, qualitative
Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL) and more
PDF, CSV
Single-mode Luminometer
Luminescence
Endpoint, kinetic
96-well microplates, white or black
Glow, Flash (Development times > 5 seconds, manual addition of reagents)
Top and bottom readout
96 Silicon Photomultiplier
100 fmol ATP/well (preliminary result)
8 decades
0.7 x 10⁻⁶ RLU
< 5 %
USB 2.0 interface with PC or WLAN
15–40 °C
9.6 x 13.8 x 4.7 cm (W x L x H)
8.6 x 12.8 x 1.7 cm (W x L x H)
Upper part 515 g
USB connection 5 V
2.5 Watts
Windows 10 (1809 or later)
macOS 11 Big Sur
Endpoint, Kinetic
Quantitative, qualitative
Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL) and more
PDF, CSV
Absorbance
Endpoint, kinetic
96-well microplates
Up to 6 LEDs
96 Photodiodes
Example combinations (FWHM 10 nm): 450, 492, 562, 620 nm/405, 450, 540, 630 nm/492, 562, 605, 650 nm. Other filter combinations in the range of 400–1000 nm possible.
0.0–4.0 OD
405 nm: ≤ 1.5 % from 0.0–2.0 OD; ≤ 3.0 % from 2.0–3.0 OD
≥ 450 nm: ≤ 1.0 % from 0.0–2.0 OD; ≤ 1.5 % from 2.0–3.0 OD
405 nm: ≤ 1.5 % + 0.010 OD from 0.0–2.0 OD; ≤ 3.0 % + 0.010 OD from 2.0–3.0 OD
≥ 450 nm: ≤ 1.0 % + 0.010 OD from 0.0–2.0 OD; ≤ 1.5 % + 0.010 OD from 2.0–3.0 OD
≤ 0.5 % + 0.005 OD from 0.0–2.0 OD; ≤ 1 % + 0.010 OD from 2.0–3.0 OD
0.001 OD
Down to 3 sec. at single wavelength
USB 2.0 interface with PC
9.6 x 15.5 x 5.7 cm (W x L x H)
8.6 x 12.8 x 1.7 cm (W x L x H)
Upper part 435 g, lower part 355 g
USB connection 5 V
2.5 Watts
Windows 10 (1809 or later)
macOS 11 Big Sur or later
Endpoint, Kinetic
Quantitative, qualitative
Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL) and more
PDF, CSV
Absorbance
Endpoint, kinetic
96-well microplates
Up to 4 LEDs
96 Photodiodes
Example combinations (FWHM 10 nm): 450, 492, 562, 620 nm/405, 450, 540, 630 nm/492, 562, 605, 650 nm. Other filter combinations in the range of 400–1000 nm possible.
0.0–4.0 OD
405 nm: ≤ 1.5 % from 0.0–2.0 OD; ≤ 3.0 % from 2.0–3.0 OD
≥ 450 nm: ≤ 1.0 % from 0.0–2.0 OD; ≤ 1.5 % from 2.0–3.0 OD
405 nm: ≤ 1.5 % + 0.010 OD from 0.0–2.0 OD; ≤ 3.0 % + 0.010 OD from 2.0–3.0 OD
≥ 450 nm: ≤ 1.0 % + 0.010 OD from 0.0–2.0 OD; ≤ 1.5 % + 0.010 OD from 2.0–3.0 OD
≤ 0.5 % + 0.005 OD from 0.0–2.0 OD; ≤ 1 % + 0.010 OD from 2.0–3.0 OD
0.001 OD
Down to 3 sec. at single wavelength
USB 2.0 interface with PC
9.6 x 15.4 x 5.5 cm (W x L x H)
900 g
USB connection 5 V
2.5 Watts
Windows 10 (1809 or later)
macOS 11 Big Sur or later
Endpoint, kinetic
Quantitative, Qualitative
Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL)
PDF, CSV
Single beam, single wavelength cuvette photometer
Absorbance
Endpoint
Micro, semi-micro and macro cuvettes
1 x LED
UV: 280, 340 nm/VIS: 400–1000 nm
0–5.0 OD
0.001 OD
≤ 0.5 % + 0.005 OD from 0.0–3.0 OD
≤ 0.5 % + 0.005 OD from 0.0–3.0 OD
≤ 0.5 % from 0.0–3.0 OD
5–50 °C
USB connection 5 V
33 mm x Ø 64 mm
150 g
Windows 10
macOS 11
Android 5