Specifications

Absorbance 96 Automate

General

Detection method

Absorbance

Detection mode

Endpoint, kinetic

Microplates types

96-well microplates

Measurement

Light source

Up to 6 LEDs

Detector

96 Photodiodes

Wavelength selection

Example combinations (FWHM 10 nm): 450, 492, 562, 620 nm/405, 450, 540, 630 nm/492, 562, 605, 650 nm. Other filter combinations in the range of 400–1000 nm possible. 

Photometric range

0.0–4.0 OD 

Linearity

405 nm: ≤ 1.5 % from 0.0–2.0 OD; ≤ 3.0 % from 2.0–3.0 OD    

≥ 450 nm: ≤ 1.0 % from 0.0–2.0 OD; ≤ 1.5 % from 2.0–3.0 OD 

Accuracy

405 nm: ≤ 1.5 % + 0.010 OD from 0.0–2.0 OD; ≤ 3.0 % + 0.010 OD from 2.0–3.0 OD

≥ 450 nm: ≤ 1.0 % + 0.010 OD from 0.0–2.0 OD; ≤ 1.5 % + 0.010 OD from 2.0–3.0 OD

Reproducibility

≤ 0.5 % + 0.005 OD from 0.0–2.0 OD; ≤ 1 % + 0.010 OD from 2.0–3.0 OD

Resolution

0.001 OD 

Read time

Down to 3 sec. at single wavelength

Data output

USB 2.0 interface with PC 

Physical characteristics

Dimensions Reader

9.6 x 15.5 x 5.7 cm (W x L x H)

Dimensions SBS Adapter

8.6 x 12.3 x 1.7 cm (W x L x H)

Weight Reader

Upper part 435 g, lower part 355 g

Power supply

USB connection 5 V

Power consumption

2.5 Watts

App

System Requirements

Microsoft Windows

Windows 7 or higher

macOS

High Sierra 10.13. or higher

App features

Measurement modes

Endpoint, Kinetic

Evaluation methods

Quantitative, qualitative

Regression methods

Point-to-Point, Linear, Four Parameter Logistic (4PL), Five Parameter Logistic (5PL) and more

Export formats

PDF, CSV

Please rotate your Device